A frameless method for stereotactic radiotherapy

Autor: J W Rieke, S S Vermeulen, Douglas L. Jones, D A Christopherson, Mark D. Hafermann, John T. Washington, J J Travaglini
Rok vydání: 1993
Předmět:
Zdroj: The British Journal of Radiology. 66:1142-1150
ISSN: 1748-880X
0007-1285
DOI: 10.1259/0007-1285-66-792-1142
Popis: A frameless method for stereotactic multiple arc radiotherapy (SMART) is described. Three short gold wires are implanted in the scalp approximately 100 mm apart. These are localized in a computed tomographic or angiographic study along with the target. Subsequently the gold markers are localized on beam films and the target position calculated using a computer program ISOLOC. This program provides the couch movements required to move the target to the isocentre and a micropositioner attached to the couch is used to make the adjustment. Beam films are repeated until the movements required are less than 1 mm in any direction. It is shown that the simple procedures of implanting the markers subcutaneously do not provide a stable reference system in about 25% of patients and the markers are now screwed into the cranium. The precision of the method is evaluated by phantom studies and measurements taken during several hundred treatments.