Pediatric Simulation of Intrinsic PEEP and Patient-Ventilator Trigger Asynchrony During Mechanical Ventilation
Autor: | Amanda J Nickel, Howard B Panitch, Joseph M McDonough, Etze Chotzoglou, Julian L Allen |
---|---|
Rok vydání: | 2022 |
Předmět: | |
Zdroj: | Respiratory care. 67(11) |
ISSN: | 1943-3654 |
Popis: | Intrinsic PEEP during mechanical ventilation occurs when there is insufficient time for expiration to functional residual capacity before the next inspiration, resulting in air trapping. Increased expiratory resistance (RUsing a Servo lung model (ASL 5000) and LTV 1200 ventilator in pressure control mode, we developed a passive model demonstrating how elevated RIntrinsic PEEP increased significantly with increasing RA passive lung model describes the development of increasing intrinsic PEEP with increasing R |
Databáze: | OpenAIRE |
Externí odkaz: |