Instrument to determine the haemoglobin-oxygen equilibrium curve based on the calculated diffusion of oxygen across a semi-permeable membrane
Autor: | David Eric Lees, William H. Schuette, L. E. Thibault |
---|---|
Rok vydání: | 1980 |
Předmět: |
Materials science
Diffusion equation Sample (material) Diffusion Biomedical Engineering Analytical chemistry chemistry.chemical_element Membranes Artificial Partial pressure Silicone rubber Oxygen Computer Science Applications chemistry.chemical_compound Hemoglobins Membrane chemistry Semipermeable membrane Blood Gas Analysis |
Zdroj: | Medicalbiological engineeringcomputing. 18(4) |
ISSN: | 0140-0118 |
Popis: | A technique is presented for automatically obtaining the haemoglobin-oxygen equilibrium curve from samples of concentrated haemoglobin solutions or whole blood The amount of oxygen exchanged to the sample across a silicone rubber membrane is calculated from the diffusion equation using a calibrated gas source and the measured pO2 of the sample. With this system it is possible to calculate the entire equilibrium curve from a set of pO2 measurements made as the desaturated sample is raised to an oxygen partial pressure of 150 mm Hg or greater if desired. |