Carbon dioxide rebreathing caused by deformed silicon leaflet in the expiratory unidirectional valve

Autor: Sandeep Kumar Mishra, Lenin Babu Elakkumanan, Arumugam Vasudevan, Satyen Parida
Rok vydání: 2013
Předmět:
Zdroj: Journal of Anaesthesiology Clinical Pharmacology, Vol 29, Iss 1, Pp 108-110 (2013)
Journal of Anaesthesiology, Clinical Pharmacology
ISSN: 0970-9185
DOI: 10.4103/0970-9185.105816
Popis: Rebreathing of carbon dioxide caused by incompetent ‘cage and disc’ unidirectional valves has been reported earlier. Some manufacturers have changed the design of unidirectional valves to ‘flexible leaflets’. We report a series of cases where a deformed membrane leaflet in expiratory unidirectional valves led to rebreathing of carbon dioxide.