A new manometry device for evaluating the sphincter of Oddi using a fiber-optic pressure sensor
Autor: | Tadao Matsunaga, Atsushi Masamune, Takashi Hanzawa, Atsushi Kanno, Tomoyuki Koike, Tooru Shimosegawa, Katsunori Iijima, Yoichi Haga |
---|---|
Rok vydání: | 2017 |
Předmět: |
medicine.medical_specialty
business.industry Bile duct Manometry Urology Equipment Design medicine.disease digestive system Pressure sensor Endoscopy Gastrointestinal 03 medical and health sciences 0302 clinical medicine medicine.anatomical_structure Pancreatitis Sphincter of Oddi Dysfunction 030220 oncology & carcinogenesis Sphincter of Oddi dysfunction Sphincter of Oddi medicine Humans 030211 gastroenterology & hepatology Surgery A fibers business |
Zdroj: | Minimally invasive therapyallied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy. 27(4) |
ISSN: | 1365-2931 |
Popis: | Introduction: Although measuring the pressure of the sphincter of Oddi and the bile duct is considered to be an important examination, called Sphincter of Oddi manometry (SOM), some complications related to the SOM device remain unsolved.Material and methods: To decrease adverse complications, we developed a 0.46 mm manometry and we performed some in vitro studies.Results: We successfully developed a 0.46 mm SOM. The diameter is the thinnest size used in endoscopic examinations. The results of in vitro studies show the suitability as SOM.Conclusion: This device will decrease the risks related to SOM examination. To confirm the safety and feasibility, further studies including in vivo studies will be needed. |