Multi-matrix assay of cortisol, cortisone and corticosterone using a combined MEPS-HPLC procedure
Autor: | Maria Addolorata Saracino, Nadia Ghedini, Maria Augusta Raggi, Corrado Iacono, Lorenzo Somaini, Gilberto Gerra |
---|---|
Přispěvatelé: | Maria A. Saracino, Corrado Iacono, Lorenzo Somaini, Gilberto Gerra, Nadia Ghedini, Maria A. Raggi |
Rok vydání: | 2014 |
Předmět: |
Hydrocortisone
Clinical Biochemistry Pharmaceutical Science Adrenal Cortex Hormone Solid-phase microextraction High-performance liquid chromatography Analytical Chemistry Dried blood and urine spot Plasma Adrenal Cortex Hormones Limit of Detection Drug Discovery Corticosteroid Chromatography High Pressure Liquid Spectroscopy medicine.diagnostic_test Heroin Dependence Chemistry Healthy Volunteer Healthy Volunteers Analgesics Opioid Substance Abuse Detection Spectrophotometry Microextraction by packed sorbent Buffer Human medicine.drug Analyte Bioanalysis Acetonitriles Sorbent Reproducibility of Result Phosphate Buffers Phosphates medicine Humans Acetonitrile Saliva Solid Phase Microextraction Detection limit Chromatography Drug Discovery3003 Pharmaceutical Science Reproducibility of Results Cortisone Immunoassay Adsorption Corticosterone |
Zdroj: | Journal of Pharmaceutical and Biomedical Analysis. 88:643-648 |
ISSN: | 0731-7085 |
Popis: | The development and validation of a bioanalytical assay for the simultaneous determination of cortisol, cortisone and corticosterone levels in several matrices, such as saliva, plasma, blood and urine samples have been described. The method is based on a rapid test which combines a microextraction by packed sorbent procedure and liquid chromatography-diode array technique. Chromatographic separation of the analytes (cortisol, cortisone and corticosterone) and the internal standard (methylprednisolone) was achieved in less than 10 min on a reversed-phase pentafluorophenyl column using a mobile phase com-posed of phosphate buffer and acetonitrile. The assay was performed after an innovative microextraction procedure by means of C8 sorbent which guaranteed good clean-up of the matrices and satisfactory extraction yield of the analytes. Moreover, the method gave linear results over a range of 5–100 ng mL−1 and showed good selectivity and precision. This method was successfully applied for quantifying corticosteroids in specific matrices derived from some healthy volunteers in comparison to two socially diversified groups, namely former heroin addicts undergoing opioid replacement therapy and poly-drug abusers The development and validation of a bioanalytical assay for the simultaneous determination of cortisol, cortisone and corticosterone levels in several matrices, such as saliva, plasma, blood and urine samples have been described. The method is based on a rapid test which combines a microextraction by packed sorbent procedure and liquid chromatography-diode array technique. Chromatographic separation of the analytes (cortisol, cortisone and corticosterone) and the internal standard (methylprednisolone) was achieved in less than 10min on a reversed-phase pentafluorophenyl column using a mobile phase composed of phosphate buffer and acetonitrile. The assay was performed after an innovative microextraction procedure by means of C8 sorbent which guaranteed good clean-up of the matrices and satisfactory extraction yield of the analytes. Moreover, the method gave linear results over a range of 5-100ngmL-1 and showed good selectivity and precision.This method was successfully applied for quantifying corticosteroids in specific matrices derived from some healthy volunteers in comparison to two socially diversified groups, namely former heroin addicts undergoing opioid replacement therapy and poly-drug abusers. © 2013 Elsevier B.V. |
Databáze: | OpenAIRE |
Externí odkaz: |