Flow-injection binding assays: A way to increase the speed in binding analyses
Autor: | Per Berdén, Torbjörn G.I. Ling, Bo Mattiasson |
---|---|
Rok vydání: | 1989 |
Předmět: |
Mannosides
Carbohydrates Biophysics Buffers Biochemistry Horseradish peroxidase Antibodies Photometry Absorbance Concanavalin A Antigens Molecular Biology Chromatography High Pressure Liquid Horseradish Peroxidase Chromatography biology Chemistry Immunochemistry Ligand binding assay Lectin Cell Biology Enzyme assay Molecular Weight biology.protein Indicators and Reagents Gels Protein Binding Peroxidase |
Zdroj: | Analytical Biochemistry. 181:379-382 |
ISSN: | 0003-2697 |
Popis: | A flow-injection analysis system was equipped with a small column containing immobilized concanavalin A. Pulses containing glucosides or glycoproteins were passed over the column, the lectin bound the carbohydrates. By using horseradish peroxidase as a labeled carbohydrate and letting it compete with other glucosides or mannosides a competitive binding assay for the latter was set up. When the enzyme activity had been evaluated, the column was rinsed and reconditioned, allowing a new assay to be run. To speed up the assay, substrates for the enzyme marker, peroxidase, were present in the perfusing buffer. A computerized evaluation of the absorbance peak allowed the time of the assay cycle to be reduced to 70 s. The sensitivity of this binding assay was fully comparable with those reported for other systems using the same reactants. |