MRI Artifact Masquerading as Orbital Disease
Autor: | Andrew S. Eiseman, Charles E. Swallow, Bruce A. Brown |
---|---|
Rok vydání: | 2001 |
Předmět: |
Adult
Male medicine.medical_specialty genetic structures CONTRAST ENHANCED MRI Vision Disorders Gadolinium Inversion recovery Normal MRI Tertiary care Orbital disease Orbital Diseases medicine Humans Artifact (error) medicine.diagnostic_test business.industry Magnetic resonance imaging Image Enhancement equipment and supplies Magnetic Resonance Imaging Ophthalmology Radiology Abnormality Artifacts business Nuclear medicine |
Zdroj: | International Ophthalmology. 24:343-347 |
ISSN: | 0165-5701 |
DOI: | 10.1023/b:inte.0000006832.22117.f6 |
Popis: | Objectives: To describe the diagnostic challenge associated with magnetic resonance imaging (MRI) artifact and to describe methods of minimizing this artifact. Methods: Gadolinium enhancement was demonstrated to be artifact by using an inversion recovery pulse sequence technique. Results: A patient with complaints of painless loss of vision also had radiologic abnormalities that led to subsequent referral to a tertiary care facility. Increased signals on the fat-suppressed contrast enhanced MRI was demonstrated to be artifact by using an inversion recovery technique which produced a normal MRI of the orbit. Conclusions: Clinical confusion leading to unnecessary and costly evaluation may be avoided if one considers the possibility of artifact when reviewing magnetic resonance studies. When magnetic susceptibility is suspected, the use of an inversion recovery sequence can better delineate the true nature of the abnormality. |