Membrane-tube-type glaucoma shunt device for refractory glaucoma surgery
Autor: | Jong Chul Han, Byung Heon Ahn, Young Hoon Hwang |
---|---|
Rok vydání: | 2016 |
Předmět: |
Adult
Male medicine.medical_specialty Intraocular pressure Time Factors genetic structures medicine.medical_treatment Visual Acuity Glaucoma Prosthesis Design Prosthesis Implantation 03 medical and health sciences Cellular and Molecular Neuroscience 0302 clinical medicine Refractory Ophthalmology medicine Glaucoma surgery Humans Trabeculectomy Tube (fluid conveyance) Glaucoma Drainage Implants Aged Retrospective Studies Aged 80 and over business.industry Stent Capsule Middle Aged medicine.disease eye diseases Sensory Systems Surgery Treatment Outcome 030221 ophthalmology & optometry Female sense organs business Sclera 030217 neurology & neurosurgery Follow-Up Studies |
Zdroj: | Graefe's Archive for Clinical and Experimental Ophthalmology. 255:163-169 |
ISSN: | 1435-702X 0721-832X |
DOI: | 10.1007/s00417-016-3510-z |
Popis: | To evaluate the safety and efficacy of a novel membrane-tube (MT)-type glaucoma shunt device for refractory glaucoma surgery. The device consists of an expanded polytetrafluoroethylene membranous reservoir, as well as a silicone tube (300-μm external and 200-μm internal diameter) with an intraluminal stent. We named the device “Finetube MT”. The Finetube MT was implanted into 44 glaucomatous eyes that had insufficient intraocular pressure (IOP) control despite medical treatment or previous trabeculectomy. The membranous reservoir was placed underneath the Tenon’s capsule, with each end located below the recti muscles; the tube was placed in the anterior chamber through a partial-thickness scleral track. We investigated the baseline and post-operative IOP values, the number of IOP-lowering medications used, and complications. The mean age of the subjects was 51.6 ± 17.2 years, and the mean follow-up duration was 22.5 ± 12.0 months. One year after the surgery, the mean IOP had decreased from 32.8 ± 12.2 mmHg to 16.9 ± 6.4 mmHg (48.5 % reduction; p |
Databáze: | OpenAIRE |
Externí odkaz: |