Measurement of the anterior chamber depth and ocular axial length: transpalpebral B-mode ultrasound using an 18-MHz linear probe compared with the IOL Master 500
Autor: | Juan José Criado-Álvarez, Juan José Arévalo-Manso, Juan Jesus Gomez-Herrera, Maria Satrustegui-Lapetra, Lorenzo Ismael Perez-Sanchez, Julia Gutierrez-Vazquez |
---|---|
Rok vydání: | 2021 |
Předmět: |
Biometry
Acoustics and Ultrasonics Radiological and Ultrasound Technology Anterior Chamber business.industry B mode ultrasound Limits of agreement Ultrasound Reproducibility of Results Axial length Conventional ultrasound Mean difference Cross-Sectional Studies Ocular axial length Humans Medicine Radiology Nuclear Medicine and imaging Linear probe business Nuclear medicine Ultrasonography |
Zdroj: | Medical Ultrasonography. 23:48 |
ISSN: | 2066-8643 1844-4172 |
Popis: | Aims: To determine the reliability of transpalpebral ultrasound in B-mode (B-TUS) with a high-resolution linear probe (18 MHz) in estimating both the ocular anterior chamber depth (ACD) and axial length (AL), as well as its agreement with the IOL Master 500 optical biometer. Material and methods : Cross-sectional study on 82 eyes of 41 volunteers with no history of eye disease. ACD and AL were determined using B-TUS and the IOL Master 500. The agreement between the two tech - niques and the variability of B-TUS (inter- and intra-observer) were analyzed with the Bland-Altman method. To this end, the mean difference between measures±1.96 SD was calculated to determine the limits of agreement (LoA). Results: The mean difference±1.96SD between B-TUS and the IOL Master 500 was -0.41±0.25mm for ACD (p |
Databáze: | OpenAIRE |