Intraocular lens power calculation formula accuracy: Comparison of 12 formulas for a trifocal hydrophilic intraocular lens
Autor: | Jorge Peraza-Nieves, David Flikier, José-María Sánchez-González, Elvira Colmenero-Reina, Francisco-Javier Castro-Alonso, Jose-Luis García-Madrona, Alvaro Rodriguez-Raton, Carlos Rocha-de-Lossada |
---|---|
Rok vydání: | 2020 |
Předmět: |
Male
0301 basic medicine Optics and Photonics Biometry medicine.medical_treatment Emmetropia Basis function Intraocular lens Refraction Ocular 03 medical and health sciences 0302 clinical medicine Optics Lens Implantation Intraocular medicine Humans Retrospective Studies Lenses Intraocular Physics Phacoemulsification business.industry General Medicine Axial Length Eye Ophthalmology 030104 developmental biology Intraocular lens power 030221 ophthalmology & optometry Female Intraocular lens power calculation business |
Zdroj: | European Journal of Ophthalmology. 31:2981-2988 |
ISSN: | 1724-6016 1120-6721 |
DOI: | 10.1177/1120672120980690 |
Popis: | Purpose: To evaluate the accuracy of 12 intraocular lens (IOL) power formulas; Barrett Universal II, Emmetropia Verifying Optical (EVO), Haigis, Hill-Radial Basis Function (RBF), Hoffer Q, Holladay I, Kane, Ladas Super Formula, Olsen Lenstar, Panacea, Pearl-DGS, Sanders-Retzlaff-Kraff/theoretical (SRK/T). In addition, an analysis of the efficacy as a function of the axial length was performed. Methods: About 171 from 93 patients: 68 male eyes and 103 female eyes. Twelve IOL power formula calculations were studied with one IOL platform (trifocal hydrophilic IOL, FineVision Micro F), one biometer (Lenstar LS 900), one topographer (CSO Sirius Topographer), one surgeon, and one optometrist. Optimization were determined to be zeroed mean refractive prediction error. Mean error (ME), mean absolute error (MAE), median absolute error (MedAE) and refractive accuracy within ±1.00 D was calculated. Axial length was split in short and medium eyes. Results: One hundred and seventy eyes were included. Formulas were ranked by percentage within ±0.50 diopters and MAE (D). Among all eyes, Olsen 86.55% (0.273 D) and Barrett Universal II 86.55% (0.285D). For short eyes (Conclusion: Olsen and Barrett formula obtained excellent accuracy for overall eyes. Kane and Olsen formula obtained the best results in short eyes. For medium axial length Barrett formula achieved the best accuracy results. |
Databáze: | OpenAIRE |
Externí odkaz: |