Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation
Autor: | Claire E. Warren, Mark Anderson, Martyn W. Richards |
---|---|
Rok vydání: | 2010 |
Předmět: |
Adult
medicine.medical_specialty Time Factors Defibrillation medicine.medical_treatment Single coil Implantable defibrillator Physiology (medical) Internal medicine medicine Humans Lead (electronics) Pacing impedance medicine.diagnostic_test business.industry Middle Aged Implantable cardioverter-defibrillator Conductor Defibrillators Implantable Heart Arrest Prosthesis Failure Radiography Cardiology Tachycardia Ventricular Female Cardiology and Cardiovascular Medicine business Chest radiograph |
Zdroj: | Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 12(8) |
ISSN: | 1532-2092 |
Popis: | Two patients with the same model of single-coil active fix implantable defibrillator lead presented with evidence of lead malfunction, in one case with an abrupt rise in pacing impedance and threshold and in the other with evidence of over-sensing. In both cases, the chest radiograph showed separation of conductors in the defibrillation lead from the main body of the lead. |