Comparison of noninvasive continuous arterial waveform analysis (Nexfin) with transthoracic Doppler echocardiography for monitoring of cardiac output
Autor: | Albertus J. Voogel, Anna G.E. van der Spoel, Anja Folkers, R. Arthur Bouwman, Christa Boer |
---|---|
Přispěvatelé: | Anesthesiology, ICaR - Circulation and metabolism |
Jazyk: | angličtina |
Předmět: |
Adult
Male Cardiac function curve Cardiac output medicine.medical_specialty Brachial Artery Doppler echocardiography Clinical study Method comparison symbols.namesake Pulse waveform Internal medicine medicine Humans Aged Monitoring Physiologic Aged 80 and over medicine.diagnostic_test business.industry Arterial pulse contour analysis Reproducibility of Results Blood Pressure Determination Middle Aged Blood Pressure Monitors Echocardiography Doppler body regions Blood pressure Anesthesiology and Pain Medicine Validation studies symbols Cardiology Female Doppler ultrasound business Doppler effect Echocardiography Transesophageal |
Zdroj: | Journal of Clinical Anesthesia, 24(4), 304-309. Elsevier Inc. van der Spoel, A G E, Voogel, A J, Folkers, A, Boer, C & Bouwman, R A 2012, ' Comparison of noninvasive continuous arterial waveform analysis (Nexfin) with transthoracic Doppler echocardiography for monitoring of cardiac output ', Journal of Clinical Anesthesia, vol. 24, no. 4, pp. 304-309 . https://doi.org/10.1016/j.jclinane.2011.09.008 |
ISSN: | 0952-8180 |
DOI: | 10.1016/j.jclinane.2011.09.008 |
Popis: | Study ObjectivesTo compare the Nexfin cardiac output (CO) with the CO obtained from transthoracic Doppler echocardiography (TTE) during routine cardiac function screening.DesignObservational clinical study.SettingEchocardiography laboratory.Patients40 ASA physical status 1 and 2 patients scheduled for routine TTE examination.InterventionsNone.Measurements and Main ResultsIn 40 patients scheduled for routine TTE examination, we obtained simultaneous CO measurements with Doppler ultrasound and derived from Nexfin blood pressure measurements. Correlation and level of agreement between Nexfin and TTE were analyzed using Pearson correlation coefficient and Bland-Altman plots. The Pearson correlation coefficient for Nexfin versus TTE was 0.68 (CI: 0.46 - 0.82, P < 0.0001). Bland-Altman analysis showed a bias of 0.51 ± 1.1 L/min and limits of agreement of -1.6 to 2.6 L/min, with a percentage error of 39%.ConclusionsConsidering limits of precision of CO measurements with Doppler echocardiography (± 30%), the agreement between noninvasive CO measurement with the Nexfin and TTE is reasonable. |
Databáze: | OpenAIRE |
Externí odkaz: |