Comparison of two digital intraoral radiography imaging systems as a function of contrast resolution and exposure time

Autor: Mutlu Özcan, Kader Cesur Aydin, Oǧuzhan Demirel
Přispěvatelé: University of Zurich, Aydin, Kader C
Rok vydání: 2020
Předmět:
Zdroj: Minerva Stomatologica. 69
ISSN: 1827-174X
0026-4970
DOI: 10.23736/s0026-4970.19.04286-9
Popis: Background: To compare the image quality of two different digital imaging systems; one photostimulable phosphor plate system (PSP) and a direct digital radiography system with CMOS imaging sensor; via evaluating contrast resolution among four different exposure times. Methods: Endodontically treated incisor teeth embedded in paraffin blocks are aligned next to a 99.5% Al wedge and exposed for 0.8, 0.1,0.125 and 0.16 seconds using both the CMOS and PSP systems. Using ImageJ software, 5 isometric and isogridded ROI from each root filling area and isometric ROI from the Al stepwedge were calculated. Results: Evaluation of the total of 120 images displayed that PSP system produced significantly higher contrast resolution (P0.05). Conclusions: The contrast resolution was higher using the PSP system. It can be estimated that, filling material will be more obvious under lower doses using PSP.