Design of a resilient ring for middle ear’s chamber stapes prosthesis
Autor: | Konrad Kamieniecki, Emilia Anna Kiryk, Monika Kwacz |
---|---|
Rok vydání: | 2018 |
Předmět: |
Materials science
medicine.medical_treatment Finite Element Analysis 0206 medical engineering Biomedical Engineering Ear Middle Bioengineering 02 engineering and technology Stapes Surgery Prosthesis Design Prosthesis Prosthesis Implantation 03 medical and health sciences 0302 clinical medicine medicine Humans In patient Stapes Orthodontics General Medicine Perilymph medicine.disease 020601 biomedical engineering Computer Science Applications Footplate Human-Computer Interaction Ossicular Prosthesis medicine.anatomical_structure Middle ear Ligament Computer-Aided Design Otosclerosis 030217 neurology & neurosurgery |
Zdroj: | Computer Methods in Biomechanics and Biomedical Engineering. 21:771-779 |
ISSN: | 1476-8259 1025-5842 |
DOI: | 10.1080/10255842.2018.1519070 |
Popis: | This paper presents the process of designing a new elastic element replacing a membrane in the chamber stapes prosthesis (ChSP). The results of the study are volume displacement characteristics obtained for the prosthesis and physiological stapes. Simulation tests on a 3D CAD model have confirmed that a properly designed ring can stimulate perilymph with the same or greater efficacy as the physiological stapes footplate placed on the elastic annular ligament. The ChSP with a new elastic element creates a good chance of improving hearing in patients suffering from otosclerosis. |
Databáze: | OpenAIRE |