Beat-to-beat detection of fetal heart rate: Doppler ultrasound cardiotocography compared to direct ECG cardiotocography in time and frequency domain

Autor: Barbara Vermeulen, S. Guid Oei, Edith D M ten Broeke, Peter Andriessen, Pieter F. F. Wijn, Ralph C M Berendsen, Chris H.L. Peters
Přispěvatelé: Transport in Permeable Media, Signal Processing Systems, Biomedical Diagnostics Lab
Rok vydání: 2004
Předmět:
Zdroj: Physiological Measurement, 25(2), 585-593. Institute of Physics
ISSN: 0967-3334
Popis: In order to obtain power spectral information on the fetal heart rate in stages of pregnancy earlier than labor an algorithm has been developed to calculate the fetal heart rate on a beat-to-beat basis from Doppler ultrasound cardiotocographic signals. The algorithm was evaluated by comparing the calculated fetal heart rate with the heart rate determined from direct ECG signals measured with a scalp electrode. Heart rates were compared both in time and frequency domain. In the time domain the results achieved by both methods correlate well (correlation coefficient = 0.977 (p