The Use of Digitized Port Films in Assessing New Treatment Modalities

Autor: J.A. Berry, J.E. Aldrich, J.W. Andrew
Rok vydání: 1990
Předmět:
Zdroj: Medical Dosimetry. 15:121-125
ISSN: 0958-3947
DOI: 10.1016/0958-3947(90)90049-n
Popis: An inexpensive computer-based imaging system has been used to compare simulator and portal radiographs. The films were digitized, normalized, and then overlaid for measurement. In a test group of patients, a new treatment protocol for pelvis irradiation was evaluated using this technique. For 90% of the patients, the average error was found to be smaller than 5 mm, and there were no field placement errors greater than 1 cm.
Databáze: OpenAIRE