Broken inspiratory valve detected by capnographic waveforms during general anesthesia: A case report

Autor: Mi-Young Kim, Byung Woo Min, Jong Suk Ban, Sang Kon Lee, Ji Hyang Lee, Eun Joo Kim
Rok vydání: 2019
Předmět:
Zdroj: Korean journal of anesthesiology. 57(4)
ISSN: 2005-7563
Popis: Unidirectional valve malfunction causes re-breathing of expired gas during anesthesia. Capnography is a useful method for monitoring the integrity of anesthetic equipment such as the unidirectional valves in a circular system. We report, here a case in which the capnography did not sound any rebreathing alarm, but the capnogram showed a prolonged plateau, an apparently short inspiratory period and a gradually sloping descending limb that stopped just short of baseline. As a result, capnography helped to identify an inspiratory unidirectional valve malfunction during the course of anesthesia. This study emphasizes the need to analyze the capnogram during every procedure involving anesthesia, in addition to preoperative checking of the Unidirectional valve.