Development and evaluation of a method to measure wrist range of motion on paretic hand rehabilitation device

Autor: Igor Gaponov, Bong-Keun Jung, Youn-Kyung Kang, Jung-Yeon Kim, Hyun-Seok Sung
Rok vydání: 2017
Předmět:
Zdroj: ICORR
DOI: 10.1109/icorr.2017.8009434
Popis: This paper presents the development of wrist ROM measurement system that was implemented on a rehabilitation glove for paretic hand, Easy-Flex, using a rotary potentiometer. This approach enables the device to measure the angles of up to 80° with the approximate resolution of 0.1°. The method along with a universal goniometer and a smartphone clinometer application were tested on 17 healthy participants in terms of reliability. The three instruments were able to accurately measure three different angles (flexion: −10°, neutral: 0°, and extension: 20°) with measurement errors less than 2°. Moreover, ICCs of the proposed method were close to 0.90 indicating the method implemented on Easy-Flex is reliable for measuring wrist flexion and extension measurement.
Databáze: OpenAIRE