Development and evaluation of a method to measure wrist range of motion on paretic hand rehabilitation device
Autor: | Igor Gaponov, Bong-Keun Jung, Youn-Kyung Kang, Jung-Yeon Kim, Hyun-Seok Sung |
---|---|
Rok vydání: | 2017 |
Předmět: |
Male
Wrist Joint Engineering Measure (physics) Wrist Prosthesis Design 03 medical and health sciences 0302 clinical medicine medicine Humans Potentiometer Range of Motion Articular Simulation Reliability (statistics) 030222 orthopedics business.industry System of measurement Rehabilitation 030229 sport sciences Paresis medicine.anatomical_structure Goniometer Female Inclinometer Range of motion business |
Zdroj: | ICORR |
DOI: | 10.1109/icorr.2017.8009434 |
Popis: | This paper presents the development of wrist ROM measurement system that was implemented on a rehabilitation glove for paretic hand, Easy-Flex, using a rotary potentiometer. This approach enables the device to measure the angles of up to 80° with the approximate resolution of 0.1°. The method along with a universal goniometer and a smartphone clinometer application were tested on 17 healthy participants in terms of reliability. The three instruments were able to accurately measure three different angles (flexion: −10°, neutral: 0°, and extension: 20°) with measurement errors less than 2°. Moreover, ICCs of the proposed method were close to 0.90 indicating the method implemented on Easy-Flex is reliable for measuring wrist flexion and extension measurement. |
Databáze: | OpenAIRE |
Externí odkaz: |