Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube
Autor: | Sunghwan Shin, Hyung-Doo Park, Jongwon Oh |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2021 |
Předmět: |
Analyte
Barricor tube Clinical Biochemistry Thyrotropin Reference range Serum separating tube Brief Communication Bias Blood plasma Humans Tube (fluid conveyance) Centrifugation Immunoassay Clinical Chemistry Blood Specimen Collection Chromatography L-Lactate Dehydrogenase Chemistry Biochemistry (medical) fungi Becton dickinson General Medicine Turnaround time Plasma separating tube Analytes Vitamin B 12 Blood Collection Tube Vacutainer Blood Chemical Analysis |
Zdroj: | Annals of Laboratory Medicine |
ISSN: | 2234-3814 2234-3806 |
Popis: | The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of -10.29% and -9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit. |
Databáze: | OpenAIRE |
Externí odkaz: |