Intimal hyperplasia within biliary Wallstents: failure of recanalisation by insertion of a second endoprosthesis
Autor: | B J McKeown, I S Benjamin, W L Wong, A. Adam, P Jeer, B T Jackson |
---|---|
Rok vydání: | 1999 |
Předmět: |
Male
Reoperation medicine.medical_specialty Intimal hyperplasia Technical success Stent occlusion medicine Humans Radiology Nuclear Medicine and imaging cardiovascular diseases Aged Cholangiopancreatography Endoscopic Retrograde Common Bile Duct Hyperplasia medicine.diagnostic_test business.industry Follow up studies Interventional radiology General Medicine Cholestasis Extrahepatic Middle Aged medicine.disease Prosthesis Failure Surgery Female Stents Radiology business Follow-Up Studies |
Zdroj: | European Radiology. 9:630-633 |
ISSN: | 1432-1084 0938-7994 |
DOI: | 10.1007/s003300050722 |
Popis: | We report two patients with benign biliary strictures in whom we attempted recanalisation of metallic biliary endoprostheses, occluded by intimal hyperplasia, by the insertion of further endoprostheses within the occluded stents. Initial technical success was achieved in deploying the stents and restoring patency with elimination of mural filling defects. However, we found the intimal hyperplasia to be restrained for less than 48 h. From our initial results it appears that biliary metallic stent occlusion by intimal hyperplasia is not effectively treated by insertion of a second endoprosthesis. |
Databáze: | OpenAIRE |
Externí odkaz: |