Electrical parameters for physiological His–Purkinje pacing vary by implant location in an ex vivo canine model
Autor: | Zhongping Yang, Mary Lauren Mesich, Alexander R. Mattson, Paul A. Iaizzo, Elizabeth Mattson |
---|---|
Rok vydání: | 2019 |
Předmět: |
Qrs morphology
Bundle of His Pacemaker Artificial medicine.medical_specialty Action Potentials 030204 cardiovascular system & hematology Canine heart Electrocardiography 03 medical and health sciences QRS complex Dogs 0302 clinical medicine Physiology (medical) Internal medicine Atrial Fibrillation medicine Animals 030212 general & internal medicine Lead (electronics) business.industry Cardiac Pacing Artificial Disease Models Animal Cardiology Implant Cardiology and Cardiovascular Medicine business Lead Placement Canine model Ex vivo |
Zdroj: | Heart Rhythm. 16:443-450 |
ISSN: | 1547-5271 |
DOI: | 10.1016/j.hrthm.2018.09.009 |
Popis: | Background Permanent His-bundle pacing (HBP) is an attractive, perhaps more physiological, alternative to traditional right ventricular pacing. Objective The purpose of this study was to utilize direct visualization to more comprehensively understand the anatomy central to HBP, correlating electrical lead performance to implant locations along the His-bundle (HB) pathway. Methods Canine hearts (n = 5) were isolated and reanimated using Visible Heart methodologies. Medtronic 3830 SelectSecure leads were fixated where His potentials were present. The location of each implant was mapped/binned into 4 regions approximately analogous to the proximal, penetrating, and distal HB. Locational differences in HBP capture and resultant QRS morphology were assessed. Results Average HBP capture thresholds did not significantly vary with respect to implant location (1.0-ms pulse width; P = .48). The resulting QRS morphologies from HB-paced beats varied in relation to implant location. As leads were placed further distally along the HB, the ratio of paced to native QRS complex duration increased (ΔQRSpaced/ΔQRSnative ratios—region 2: 0.84 ± 0.16; region 3: 1.04 ± 0.42; region 4: 1.74 ± 0.86). Conclusion We demonstrated correlation between the anatomic locations of HBP lead placement and resultant QRS morphologies in a reanimated canine heart model. Proximal placement along the HB pathway resulted in more favorable QRS morphologies, suggesting improved selective HBP capture, with no significant increase in HBP capture thresholds. Pacing the HB in more proximal pathway locations improved the selectivity of HBP and may confer electrical and anatomic benefits relative to distal HBP. |
Databáze: | OpenAIRE |
Externí odkaz: |