Scientific Accuracy Matters
Autor: | C. Philip Larson |
---|---|
Rok vydání: | 2021 |
Předmět: |
Kidney
Human fat Biomedical Research Chromatography business.industry Pipette law.invention Partition coefficient Laboratory flask Anesthesiology and Pain Medicine medicine.anatomical_structure Erlenmeyer flask Solubility law Anesthetics Inhalation medicine Animals Humans Cattle Tissue Distribution Hemoglobin Halothane business medicine.drug |
Zdroj: | Anesthesiology. 135:724-727 |
ISSN: | 1528-1175 0003-3022 |
DOI: | 10.1097/aln.0000000000003952 |
Popis: | The Solubility of Halothane in Blood and Tissue Homogenates. By Larson CP, Eger EI, Severinghaus JW. Anesthesiology 1962; 23:349–55. Measured samples of human and bovine blood, human hemoglobin, and tissue homogenates from human fat and both human and bovine liver, kidney, muscle, whole brain, and separated gray and white cortex were added to stoppered 2,000-ml Erlenmeyer flasks. To each flask, 0.1 ml of liquid halothane was added under negative pressure using a calibrated micropipette. After the flask was agitated for 2 to 4 h to achieve equilibrium between the gas and blood or tissue contents, a calibrated infrared halothane analyzer was used to measure the concentration of halothane vapor. Calculated partition coefficients ranged from 0.7 for water to 2.3 for blood and from 3.5 for human or bovine kidney to 6 for human whole brain or liver and 8 for human muscle. Human peritoneal fat had a value of 138. The human blood–gas partition coefficient of 2.3 as determined by this equilibration method was well below the previously published value of 3.6. |