Popis: |
The present study is aimed at exploring different scanning parameters, detectors and their orientations for time-efficient and accurate commissioning of a 6 MV clinical linear accelerator (LINAC). Beam profiles and percentage depth dose (PDD) curves were measured with a PTW dosimetry diode, a PTW Semiflex and a PinPoint ion chamber in different orientations. To acquire beam data, equidistant (step size of 0.5 mm, 1 mm, 2 mm and 3 mm) and fanline (step size of 2-0.5 mm, 2-1 mm, 3-0.5 mm and 3-1 mm) scanning modes were employed and data measurement time was recorded. Scan time per measurement point was also varied (0.2 s, 0.5 s and 1.0 s) to investigate its effect on the accuracy and acquisition time of beam data. Accuracy of the measured data was analyzed on the basis of the variation between measured data and data modeled by a treatment planning system. Beam profiles (particularly in penumbra region) were found to be sensitive to variation in scanning resolution and showed an improved accuracy with decrease in step size, while PDD curves were affected negligibly. The accuracy of beam data obtained with the PTW dosimetry diode and the PinPoint ion chamber was higher than those obtained with the PTW Semiflex ion chamber for small fields (2 × 2 cm |