A comparative study of wear laws for soft-on-hard hip implants using a mathematical wear model

Autor: F. Di Puccio, Lorenza Mattei, Enrico Ciulli
Rok vydání: 2013
Předmět:
Zdroj: Tribology International. 63:66-77
ISSN: 0301-679X
DOI: 10.1016/j.triboint.2012.03.002
Popis: Wear of UHMWPE acetabular components is the most serious threat to the long-term success of hip replacements. Consequently numerical and experimental wear simulations are of great interest. The present study proposes a mathematical wear model and compares the most recent wear laws, based on the so-called cross-shear (CS) effect. Simulations highlighted the variability of wear predictions with the wear factors/laws. Moreover a sensitivity analysis underlined that the kinematic conditions affect volumetric/linear wear more than the loading ones. This study confirms the importance of the CS in wear predictions even if some critical issues are still open, requiring further investigations.
Databáze: OpenAIRE