Plateletpheresis efficiency: a comparison of the spectra LRS and AMICUS separators

Autor: B. Barrett, P. Rojas, S. Broughton, W.H. Churchill, C. Burgio, S. Christmas, J. Neal, Richard J. Benjamin
Rok vydání: 1999
Předmět:
Zdroj: Transfusion. 39(8)
ISSN: 0041-1132
Popis: BACKGROUND: Optimizing the yields of platelet-pheresis by increasing collection efficiencies may provide several benefits: to patients, by increasing the dose in single-donor platelet (SDP) units; to the collection center, by increasing the percentage of components that may be split into double units; and/or to the donor, by reducing the duration of donation. STUDY DESIGN AND METHODS: A prospective, randomized study was undertaken to compare the efficiency of platelet collection in paired donations by 21 donors on two cell separators (Spectra LRS, version [v] 5.1; and AMICUS, v2.37). The order of donation was randomly assigned; donations were performed at least 2 weeks apart. A fixed blood volume (4000 mL) was processed by utilizing standard protocols. Findings were confirmed in a retrospective, matched study that compared the two separators by using fixed collection times (90 min). RESULTS: The AMICUS and Spectra LRS separators consistently produced white cell-reduced (