Four Assay Designs and On-Chip Calibration: Gadgets for a Sepsis Protein Array

Autor: Patricia Buchegger, Claudia Preininger
Rok vydání: 2014
Předmět:
Zdroj: Analytical Chemistry. 86:3174-3180
ISSN: 1520-6882
0003-2700
Popis: A protein microarray for the early stage diagnosis of sepsis that allows the simultaneous detection of C-reactive protein (CRP) (2-200 μg/mL), procalcitonin (PCT) (0.2-50 ng/mL), and interleukin 6 (IL-6) (2-2000 pg/mL) has been developed. To enable the parallel detection of the differently abundant analytes, the low binding affinity between CRP and phosphocholine is exploited in a "low-sensitive" sandwich assay for CRP. The calibration is integrated directly on the chip resulting in a "one patient-one array" format, to provide a user-friendly and rapid diagnostic tool. Four different assay designs are introduced: (I) the classical assay that works with biotin-streptavidin chemistry, (II) the rapid assay that is performed in a single detection step, and two ultrasensitive assay designs accomplished either by (III) an enzymatic or (IV) an antibody mediated amplification resulting in high density labeling. The assay designs were evaluated by the repetitive measurement of low, medium, and high concentration levels of commercially available certified control sera. The precision was similar across all assay designs (coefficient of variation (CV), CVintra: 8-14%; CVinter: 18-34%), while the sensitivity (limits of detection (LODs)) increased by 1 order of magnitude for the ultrasensitive assays (III, IV) and the accuracy was analyte dependent but best for the classical (I) and the antibody amplified (IV) assays.
Databáze: OpenAIRE