Validation of Central Pressure Estimation in Patients with an Aortic Aneurysm Before and After Endovascular Repair

Autor: Jenske J.M. Vermeulen, Suzanne Holewijn, Michel M.P.J. Reijnen, M. van Helvert, L. van de Velde
Přispěvatelé: Multi-Modality Medical Imaging, TechMed Centre
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Cardiovascular engineering and technology, 13, 265-278. Springer
ISSN: 1869-408X
Popis: Purpose: The aim of this study was to investigate if non-invasive central pressure estimations are accurate in patients with an abdominal aortic aneurysm, before and after endovascular repair. Secondary evaluation was if measurement-accuracy was dependent on anatomical characteristics. Methods: Procedural invasive and non-invasive pressure-measurements were performed simultaneously both before and after endovascular repair in 20 patients with an infrarenal abdominal aortic aneurysm. Invasive catheter measurements were performed in the abdominal aorta. A tonometric device was used to perform non-invasive pressure–wave-analysis at the radial artery. A generalized transfer-function was used to generate an ascending aortic waveform for both measurements, allowing for direct comparison. Results: Pre-treatment the mean differences between methods were − 5.5 mmHg (p = .904), − 11.8 (p