Polarimeter design for a nitrogen laser source

Autor: Roger Stephens
Rok vydání: 1992
Předmět:
Zdroj: Talanta. 39:1511-1516
ISSN: 0039-9140
Popis: The system uses a Brewster angle reflector as a passive amplifier. This device reduces intensity without causing a serious reduction in either the signal of interest or the signal-to-noise ratio. It also reduces the need for a high quality analyser. However, polariser quality is critical and determines the background light level at the detector. The present work allowed shot-noise-limited operation to be attained. However full optimization was not achieved because of excessive background transmission at optimum alignment.
Databáze: OpenAIRE