Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs
Autor: | Glaucia Prada Kanashiro, Giovana José Garcia Estanho, Claudia Lizandra Ricci, Felipe Franco Nascimento, João Victor Goulart Consoni Passareli, Silvia Franco Andrade, Rogério Giuffrida |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2021 |
Předmět: |
Applanation tonometry
medicine.medical_specialty Intraocular pressure dogs genetic structures direct manometry Veterinary medicine Glaucoma Tono pen SF1-1100 Ophthalmology SF600-1100 Medicine Outpatient clinic goldmann tonometry General Veterinary business.industry REBOUND TONOMETRY medicine.disease eye diseases Animal culture Goldmann tonometry applanation tonometry rebound tonometry sense organs business Research Article |
Zdroj: | Veterinary World, Vol 14, Iss 9, Pp 2444-2451 (2021) Veterinary World |
ISSN: | 2231-0916 0972-8988 |
Popis: | Background and Aim: Tonometers are an important instrument for measuring intraocular pressure (IOP) in the diagnosis of glaucoma or uveitis. This study aimed to compare the accuracy of the main types of tonometers with different IOP measurement methodologies in dogs: TonoVet and TonoVet Plus (rebound), Tono-Pen Avia Vet (applanation), and Kowa HA-2 (Goldmann applanation). Materials and Methods: IOP was measured in 152 eyes of 76 dogs. A postmortem study was performed by comparing manometry and tonometry values and calculating the correlation coefficient (r2), in vivo real IOP (manometry) among the tonometers was compared, and an outpatient study was conducted with healthy eyes and eyes with signs of glaucoma and uveitis. Results: In the postmortem study, the values of r2 in descending order were Kowa (0.989), TonoVet Plus (0.984), TonoVet (0.981), and Tono-Pen Avia Vet (0.847). The IOP values in mmHg in the in vivo study were as follows: Aneroid manometer (16.8±2.5.7), TonoVet (18.1±2.9), TonoVet Plus (20.6±2.3), Tono-Pen Avia Vet (17.1±2.5), and Kowa (16.1±1.7); in outpatient clinics: TonoVet (16.8±3.8), TonoVet Plus (19.2±2.9), Tono-Pen Avia Vet (16.2±2.4), and Kowa (15.0±1.3); glaucoma: TonoVet (30.2±3.5), TonoVet Plus (35.0±6.1), Tono-Pen Avia Vet (29.5±4.2), and Kowa (23.9±5.0); and uveitis: TonoVet (14.2±1.4), TonoVet Plus (17.6±1.9), Tono-Pen Avia Vet (13.7±2.1), and Kowa (12.6±1.7). Conclusion: There was a strong correlation between IOP values and manometry in all the tonometers. The highest values were obtained with TonoVet Plus and the lowest with Kowa HA-2. All tonometers accurately measured IOP in dogs, including the latest TonoVet Plus, which showed an excellent correlation coefficient. |
Databáze: | OpenAIRE |
Externí odkaz: |