The Use of Ion Chromatography for the Determination of Clean-In-Place-200 (CIP-200) Detergent Traces
Autor: | Wilfredo Resto, Joan Roque, Rosamil Rey, Héctor Colón, Josέ Zayas |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2006 |
Předmět: |
ion chromatography
lcsh:QD71-142 010401 analytical chemistry lcsh:Analytical chemistry CIP detergent 02 engineering and technology cleaning validation 021001 nanoscience & nanotechnology phosphate analysis 01 natural sciences Biochemistry 6. Clean water 0104 chemical sciences Analytical Chemistry Medical Laboratory Technology 0210 nano-technology Original Research |
Zdroj: | Analytical Chemistry Insights, Vol 1, Pp 5-12 (2006) Analytical Chemistry Insights |
ISSN: | 1177-3901 |
Popis: | Anion chromatography with conductivity detection was chosen as the analytical technique for the development of a cleaning validation method for clean-in-place (CIP) detergents. The method was developed and validated for the determination of traces of the detergent CIP-200. It was shown to be linear with a squared correlation coefficient (r2) of 0.9999 and the accuracy experiments presented average recoveries of 88.2% (area response factor) from stainless steel surfaces. The repeatability was found to be 1.6% and an intermediate precision of 1.9% across the range. The method was also shown to be sensitive with an average Detection Limit (DL) of 0.23 ppm and a Quantitation Limit (QL) of 0.70 ppm based on the amount of phosphate in the detergent sample. The phosphate signal was well resolved from typical ions encountered in water samples or any other interference presented from swabs and surfaces. The method was applied to cleaning validation samples and proved to be suitable for rapid and reliable quality control. |
Databáze: | OpenAIRE |
Externí odkaz: |