Lens Opacities Classification System III–based artificial intelligence program for automatic cataract grading
Autor: | Keke Zhang, Yinglei Zhang, Xixi He, Qiang Lu, Xiangjia Zhu, Xianfang Rong, Jinrui Wang, Ling Wei, Zhennan Zhao, Wenwen He, Dayong Ding, Yi Lu, Jun Wu, Lei Cai |
---|---|
Rok vydání: | 2022 |
Předmět: |
China
business.industry Computer science Mean squared prediction error Cortical cataract Cataract Sensory Systems Residual neural network Ophthalmology Artificial Intelligence Region of interest Area Under Curve Humans Surgery Artificial intelligence Posterior subcapsular cataract business Grading (tumors) |
Zdroj: | Journal of Cataract and Refractive Surgery. 48:528-534 |
ISSN: | 1873-4502 0886-3350 |
Popis: | PURPOSE To establish and validate an artificial intelligence (AI)-assisted automatic cataract grading program based on the Lens Opacities Classification System III (LOCSIII). SETTING Eye and Ear, Nose, and Throat (EENT) Hospital, Fudan University, Shanghai, China. DESIGN AI training. METHODS Advanced deep-learning algorithms, including Faster R-CNN and ResNet, were applied to the localization and analysis of the region of interest. An internal dataset from the EENT Hospital of Fudan University and an external dataset from the Pujiang Eye Study were used for AI training, validation, and testing. The datasets were automatically labeled on the AI platform in terms of the capture mode and cataract grading based on the LOCSIII system. RESULTS The AI program showed reliable capture mode recognition, grading, and referral capability for nuclear and cortical cataract grading. In the internal and external datasets, 99.4% and 100% of automatic nuclear grading, respectively, had an absolute prediction error of ≤ 1.0, with a satisfactory referral capability (area under the curve [AUC]: 0.983 for the internal dataset; 0.977 for the external dataset). 75.0% (internal dataset) and 93.5% (external dataset) of the automatic cortical grades had an absolute prediction error of ≤ 1.0, with AUCs of 0.855 and 0.795 for referral, respectively. Good consistency was observed between automatic and manual grading when both nuclear and cortical cataracts were evaluated. However, automatic grading of posterior subcapsular cataracts was impractical. CONCLUSIONS The AI program proposed in this study shows robust grading and diagnostic performance for both nuclear and cortical cataracts, based on LOCSIII. |
Databáze: | OpenAIRE |
Externí odkaz: |