The microINR portable coagulometer:analytical quality and user-friendliness of a PT (INR) point-of-care instrument

Autor: Pia Bükmann Larsen, Åsne Bakke, Jesper Revsholm, Elin Storjord, Kristin Husby, Grete Monsen, Tone Bukve, Rie McGrail, Mikael Christensen, Aslaug Johanne Risøy, Unni Marie Skålsvik, Heidi Strand, Joakim Eikeland, Jan Nybo, Elvar Theodorsson, Mette Fogh Møller, Solveig Meier Mikaelsen, Vegar Engeland Haugen, Reyes Serrano Teruel
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Pediatrics
medicine.medical_specialty
Point-of-Care Systems
Point-of-care testing
Clinical Biochemistry
Medical laboratory
User friendliness
030204 cardiovascular system & hematology
03 medical and health sciences
0302 clinical medicine
medicine
PROTHROMBIN TIMES
Humans
Outpatient clinic
International Normalized Ratio
030212 general & internal medicine
OWREN
International normalized ratio
quality control
Blood Coagulation
Point of care
Automation
Laboratory

Prothrombin time
CALIBRATION
Analysis of Variance
medicine.diagnostic_test
Norway
business.industry
Warfarin
Anticoagulants
Reproducibility of Results
General Medicine
Laboratories
Hospital

University hospital
warfarin
point-of-care testing
Emergency medicine
Prothrombin Time
business
ANTICOAGULANT-THERAPY
medicine.drug
point-of-care systems
Zdroj: Larsen, P B, Storjord, E, Bakke, Å, Bukve, T, Christensen, M, Eikeland, J, Haugen, V E, Husby, K, McGrail, R, Mikaelsen, S M, Monsen, G, Møller, M F, Nybo, J, Revsholm, J, Risøy, A J, Skålsvik, U M, Strand, H, Teruel, R S & Theodorsson, E 2017, ' The microINR portable coagulometer : analytical quality and user-friendliness of a PT (INR) point-of-care instrument ', Scandinavian Journal of Clinical & Laboratory Investigation, vol. 77, no. 2, pp. 115-121 . https://doi.org/10.1080/00365513.2016.1277433
Larsen, P B, Storjord, E, Bakke, A, Bukve, T, Christensen, M, Eikeland, J, Haugen, V E, Husby, K, McGrail, R, Mikaelsen, S M, Monsen, G, Moller, M F, Nybo, J, Revsholm, J, Risoy, A J, Skalsvik, U M, Strand, H, Teruel, R S & Theodorsson, E 2017, ' The microINR portable coagulometer : analytical quality and user-friendliness of a PT (INR) point-of-care instrument ', Scandinavian Journal of Clinical and Laboratory Investigation. Supplement, vol. 77, no. 2, pp. 115-121 . https://doi.org/10.1080/00365513.2016.1277433
DOI: 10.1080/00365513.2016.1277433
Popis: Regular measurement of prothrombin time as an international normalized ratio PT (INR) is mandatory for optimal and safe use of warfarin. Scandinavian evaluation of laboratory equipment for primary health care (SKUP) evaluated the microINR portable coagulometer (microINR(®)) (iLine Microsystems S.L., Spain) for measurement of PT (INR). Analytical quality and user-friendliness were evaluated under optimal conditions at an accredited hospital laboratory and at two primary health care centres (PHCCs). Patients were recruited at the outpatient clinic of the Laboratory of Medical Biochemistry, St Olav's University Hospital, Trondheim, Norway (n = 98) and from two PHCCs (n = 88). Venous blood samples were analyzed under optimal conditions on the STA-R(®)Evolution with STA-SPA + reagent (Stago, France) (Owren method), and the results were compared to capillary measurements on the microINR(®). The imprecision of the microINR(®) was 6% (90% CI: 5.3-7.0%) and 6.3% (90% CI: 5.1-8.3) in the outpatient clinic and PHCC2, respectively for INR ≥2.5. The microINR(®) did not meet the SKUP quality requirement for imprecision ≤5.0%. For INR
Databáze: OpenAIRE