Design of flow chamber with electronic cell volume capability and light detection optics for multilaser flow cytometry

Autor: M. A. MacCollum, W. H. Schuette, Stanley E. Shackney, H. W. Tipton, F. A. Plowman, C. A. Smith
Rok vydání: 1984
Předmět:
Zdroj: Cytometry. 5:652-656
ISSN: 1097-0320
0196-4763
DOI: 10.1002/cyto.990050616
Popis: A multibeam optical detection system has been developed with a high optical efficiency, achieved through a reduction in the number of optical interfaces employed in the system. This reduction is made possible by a combination of employing simple lenses, gluing the objective lens directly upon the face of the flow cuvette and the extraction of only one fluorescence signal from each laser beam. A modified flow chamber is also described that includes fluidic resistance elements for the elimination of most of the electric shielding normally associated with electronic cell volume measurements.
Databáze: OpenAIRE