A musculoskeletal Finite Element Study of a Unique and Customised Jaw Joint Prosthesis for the Asian Populace
Autor: | Mehak Sharma, Manoj Soni |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2020 |
Předmět: |
Orthodontics
musculoskeletal diseases Computer science medicine.medical_treatment Finite Element Analysis Joint prosthesis Asian population Management Monitoring Policy and Law Prosthesis Finite element method Finite element study Surfaces Coatings and Films Electronic Optical and Magnetic Materials stomatognathic diseases human jaw stomatognathic system Biomedical devices computational analysis Ceramics and Composites medicine Computational analysis prosthesis |
Zdroj: | Evergreen. 7(3):351-358 |
ISSN: | 2189-0420 |
Popis: | The stress concentration for a unique temporomandibular joint prosthesis was evaluated using finite element analysis in the present work. This study intended to validate temporomandibular joint prosthesis design specific to Asian population. Total joint replacement is a complicated procedure and surgery can cause numerous complications such as hardware failure, cleft, contamination or facial immobility. Hence, it is imperative to study stresses acting during jaw movement. A three-dimensional static structural finite element approach was followed. The study reflects on even stress distribution, with least detrimental effects to human natural joint. Thus, the design efficiency is validated and also helped in analysing medical ramifications for actual temporomandibular joint surgery. |
Databáze: | OpenAIRE |
Externí odkaz: |