Marfan syndrome medicine.medical_specialty Weakness genetic structures medicine.medical_treatment Cataract Extraction Article Marfan Syndrome 03 medical and health sciences 0302 clinical medicine medicine Humans Child Capsulorhexis Anterior capsulotomy Fixation (histology) business.industry Lens Subluxation Cataract surgery medicine.disease eye diseases Sensory Systems nervous system diseases Surgery Ophthalmology Capsular bag 030221 ophthalmology & optometry sense organs medicine.symptom business 030217 neurology & neurosurgery Capsular tension segment |
Zdroj: | Journal of Cataract and Refractive Surgery. 43:590-592 |
ISSN: | 0886-3350 |
DOI: | 10.1016/j.jcrs.2017.04.021 |
Popis: | We describe clinical applications and surgical techniques for a new type of capsular tension segment (CTS) for use during cataract surgery. The Ambati CTS is distinguished from other CTS devices by having 2 eyelets close to each other, which allows it to distribute tension to 2 points, avoiding too much stress at a single point on the anterior capsulotomy, which prevents peaking of the capsulorhexis, and potentially reducing the risk for anterior capsule tear. Two of these CTS devices could possibly be used to provide 4-point fixation of a capsular bag in eyes with near-complete zonular instability. We describe 4 cases and 2 surgical techniques for implanting the new CTS, 1 technique in an adult patient with zonular weakness secondary to trauma and the other in 3 children with subluxated lenses due to Marfan syndrome. |