Real Time Intraocular Pressure and Lamellar Corneal Flap Thickness in Keratomileusis
Autor: | Peter J. McDonnell, Ngamjit Kasetsuwan, Luciane L. Moreira, Sujal Shah, Steven C. Schallhorn, Daniel Sanchez DiMartino, Rey T. Pangilinan |
---|---|
Rok vydání: | 2001 |
Předmět: |
Human cadaver
medicine.medical_specialty Intraocular pressure Time Factors Materials science Suction genetic structures medicine.medical_treatment Keratomileusis Laser In Situ Reproducibility of Results Keratomileusis In Vitro Techniques Surgical Flaps eye diseases Cornea Tonometry Ocular Ophthalmology Monitoring Intraoperative Myopia medicine Humans Lamellar structure sense organs Intraocular Pressure |
Zdroj: | Cornea. 20:41-44 |
ISSN: | 0277-3740 |
DOI: | 10.1097/00003226-200101000-00008 |
Popis: | To measure real time intraocular pressure (IOP) during keratomileusis and to determine variability of corneal flap thickness by using different suction ring pressure settings.Eight human cadaver eyes, two groups of four each, were used. The suction ring of the UniversalKeratome was applied to each eye, the same as for a standard automated lamellar keratoplasty (ALK) or laser in situ keratomileusis (LASIK) procedure, to create a lamellar corneal flap. The pressure of the suction ring on each eye was raised to one of two different levels, four eyes to 488 and the remaining four to 600 mm Hg. IOP was continuously recorded by manometer, from application of the suction ring through the end of the passage of the microkeratome. Central corneal thickness was measured, both before and after each procedure.IOP increased by90 mm Hg after application of the suction ring at a pressure of either 488 mm Hg or 600 mm Hg. Corneal flaps performed under a suction ring pressure of 488 mm Hg measured 159.98+/-5 microm, whereas flap thickness for the 600-mm Hg group varied 1.4-266.2 microm. Two flaps exhibited irregular configurations.Real time IOP can be measured during keratomileusis. The results demonstrated a significant increase in IOP during the procedure. Suction ring pressure setting is an important variable in determining consistent corneal flap thickness during the keratomileusis procedure. |
Databáze: | OpenAIRE |
Externí odkaz: |