One Year Outcomes of 101 BeGraft Stent Grafts used as Bridging Stents in Fenestrated Endovascular Repairs
Autor: | Matthieu Delloye, Dominique Fabre, Jonathan Sobocinski, Rafaelle Spear, Adrien Hertault, Richard Azzauiu, Stéphan Haulon |
---|---|
Rok vydání: | 2018 |
Předmět: |
Male
medicine.medical_specialty Time Factors Bridging (networking) Databases Factual Endoleak Computed Tomography Angiography medicine.medical_treatment 030204 cardiovascular system & hematology Prosthesis Design Aortography Blood Vessel Prosthesis Implantation 03 medical and health sciences 0302 clinical medicine medicine.artery Occlusion medicine Humans Prospective Studies cardiovascular diseases 030212 general & internal medicine Renal artery Prospective cohort study Vascular Patency Covered stent Aged Ultrasonography Doppler Duplex Aortic Aneurysm Thoracic business.industry Endovascular Procedures Ultrasound Graft Occlusion Vascular Stent equipment and supplies Blood Vessel Prosthesis Surgery Treatment Outcome surgical procedures operative Female Stents Cardiology and Cardiovascular Medicine business Aortic Aneurysm Abdominal Contrast-enhanced ultrasound |
Zdroj: | European Journal of Vascular and Endovascular Surgery. 55:504-510 |
ISSN: | 1078-5884 |
DOI: | 10.1016/j.ejvs.2018.01.023 |
Popis: | Objectives To evaluate the outcomes of the second generation BeGraft balloon expandable covered stent Graft System (Bentley InnoMed, Hechingen, Germany) implanted as bridging stent grafts during fenestrated endovascular aortic repair (FEVAR) of complex aneurysms. Design This was a single centre prospective study including all consecutive patients treated by FEVAR performed with second generation BeGraft stent grafts as bridging stents. Methods Demographics of patients, diameter and length of the bridging stent grafts, technical success, re-interventions, occlusions, post-operative events, and imaging (Cone Beam CT and/or CT scan, and contrast enhanced ultrasound) were prospectively collected in an electronic database. Duplex ultrasound was performed before discharge and at 6 month follow-up. At 1 year, patients were evaluated clinically and by imaging (CT and ultrasound). Results Between November 2015 and September 2016, 39 consecutive patients (one woman) were treated with custom made fenestrated endografts (2–5 fenestrations) for complex aneurysms or type 1 endoleak after EVAR, using a variety of bridging stents including the BeGraft. All 101 BeGraft stent grafts were successfully delivered and deployed. There was no in hospital mortality. Early fenestration patency rate was 99% (96/97); the sole target vessel post-operative occlusion was secondary to a dissection of the renal artery distal to the stent. Complementary stenting was unsuccessful in recovering renal artery patency; bilateral renal stent occlusion was observed in the same patient on a CT scan performed 2 months after the procedure. He required post-operative dialysis. No additional renal impairment was observed. During follow-up (median 13 months [11–15]), all fenestrations stented with BeGraft stent grafts remained patent (95/97, 98%). One type 1b endoleak was detected and treated (2.6%). Conclusions BeGraft stent grafts used as bridging stents during FEVAR are associated with favourable outcomes at 1 year follow-up. Long-term follow-up is required to confirm these promising results. |
Databáze: | OpenAIRE |
Externí odkaz: |