Radiation Dose Reduction at Low Tube Voltage CCTA Based on the CNR Index
Autor: | Kazuo Awai, Tomokazu Okimoto, Yukari Yamashita, Naoyuki Imada, Yoshinori Funama, Takeshi Nakaura, Tomoyasu Satou, Takanori Masuda |
---|---|
Rok vydání: | 2018 |
Předmět: |
Adult
Male Computed Tomography Angiography Coronary Artery Disease 030204 cardiovascular system & hematology Coronary Angiography Radiation Dosage 030218 nuclear medicine & medical imaging Coronary artery disease 03 medical and health sciences 0302 clinical medicine Contrast-to-noise ratio Predictive Value of Tests Hounsfield scale Humans Medicine Radiology Nuclear Medicine and imaging Aged Retrospective Studies Aged 80 and over Receiver operating characteristic business.industry Radiation dose Coronary computed tomography angiography Low tube voltage Gold standard (test) Middle Aged medicine.disease ROC Curve Female business Nuclear medicine |
Zdroj: | Academic Radiology. 25:1298-1304 |
ISSN: | 1076-6332 |
DOI: | 10.1016/j.acra.2018.01.019 |
Popis: | We compared the radiation dose and diagnostic accuracy on 120- and 100-kVp coronary computed tomography angiography (CCTA) scans whose contrast-to-noise ratio (CNR) was the same.We studied 1311 coronary artery segments from 100 patients. For 120-kVp scans, the targeted image level was set at 25 Hounsfield units (HU). For 100-kVp scans, the targeted noise level was set at 30 HU to obtain the same CNR as at 120 kVp. We compared the CNR and the radiation dose on scans acquired at 120 and 100 kVp. Invasive coronary angiography (ICA) images were evaluated by an interventional coronary angiography specialist, and CCTA images were evaluated by a radiologist. Coronary artery disease was defined as a luminal narrowing ≧50% for ICA and CCTA. With ICA considered the gold standard, the diagnostic accuracy (sensitivity, specificity, positive predictive value, and negative predictive value) was analyzed on both 120- and 100-kVp CCTA images. We also compared the diagnostic accuracy for area under the receiver operating characteristic curve of the ICA and CCTA performed at 120 and 100 kVp. Two blinded observers visually evaluated the septal branch.The mean dose-length product was 48% lower at 100 kVp than at 120 kVp (P .01). Under the 120-kVp CCTA protocol, the area under the curve, 95% confidence interval, sensitivity, specificity, positive predictive value, and negative predictive value were 0.94%, 0.91%-0.96%, 94.0%, 93.0%, 82.3%, and 98.1%, respectively; at 100 kVp these values were 0.94%, 0.92%-0.97%, 96.1%, 92.0%, 85.2%, and 98.0%, respectively. Area under the receiver operating characteristic curve analysis revealed no significant difference in diagnostic accuracy between the two protocols (P = .87).At the same CNR, the 100-kVp CCTA protocol may help to reduce the radiation dose by approximately 50% compared to the 120-kVp protocol without degradation of diagnostic accuracy. |
Databáze: | OpenAIRE |
Externí odkaz: |