Noise figure limits for circular loop MR coils

Autor: Paul A. Bottomley, Ananda Kumar, William A. Edelstein
Rok vydání: 2009
Předmět:
Zdroj: Magnetic Resonance in Medicine. 61:1201-1209
ISSN: 1522-2594
0740-3194
DOI: 10.1002/mrm.21948
Popis: Circular loops are the most common MR detectors. Loop arrays offer improved signal-to-noise ratios (SNRs) and spatial resolution, and enable parallel imaging. As loop size decreases, loop noise increases relative to sample noise, ultimately dominating the SNR. Here, relative noise contributions from the sample and the coil are quantified by a coil noise figure (NF), NFcoil, which adds to the conventional system NF. NFcoil is determined from the ratio of unloaded-to-loaded coil quality factors Q. Losses from conductors, capacitors, solder joints, eddy currents in overlapped array coils, and the sample are measured and/or computed from 40 to 400 MHz using analytical and full-wave numerical electromagnetic analysis. The Qs are measured for round wire and tape loops tuned from 50 to 400 MHz. NFcoil is determined as a function of the radius, frequency, and number of tuning capacitors. The computed and experimental Qs and NFcoils agree within ∼10%. The NFcoil values for 3 cm-diameter wire coils are 3 dB, 1.9 dB, 0.8 dB, 0.2 dB, and 0.1 dB, at 1T, 1.5T, 3T, 7T, and 9.4T, respectively. Wire and tape perform similarly, but tape coils in arrays have substantial eddy current losses. The ability to characterize and reliably predict component- and geometry-associated coil losses is key to designing SNR-optimized loop and phased-array detectors. Magn Reson Med, 2009. © 2009 Wiley-Liss, Inc.