Choice of the Optimal Output Channels Configuration in the Gas Chromatograph Detector
Autor: | V. I. Platonov, S.S. Dostovalova, D.A. Uglanov |
---|---|
Rok vydání: | 2016 |
Předmět: |
gas chromatograph detector
Hydrogen business.industry Chemistry Physics::Instrumentation and Detectors Thermal conductivity detector Nuclear engineering output channels configuration temperature fields 010401 analytical chemistry Detector Electrical engineering chemistry.chemical_element 02 engineering and technology 021001 nanoscience & nanotechnology 01 natural sciences 0104 chemical sciences Energy(all) Ansys software High Energy Physics::Experiment Gas chromatography 0210 nano-technology business Gas chromatography ion detector |
Zdroj: | Energy Procedia. 100:434-438 |
ISSN: | 1876-6102 |
DOI: | 10.1016/j.egypro.2016.10.198 |
Popis: | One of the most important chromatograph elements is a detector. This article describes the way to increase the sensibility of thermal conductivity detector - the most frequently used detector - that is being developed by Chemistry Department in cooperation with other departments at Samara University. The purpose of the research is choice of the optimal output channels configuration in the gas chromatograph detector that provides uniform temperature fields distribution at constant gas-flow rate. Two options of the output channels configuration were proposed. The temperature fields in gas chromatograph detector were calculated by ANSYS software package for mixtures with various concentrations of air and hydrogen. |
Databáze: | OpenAIRE |
Externí odkaz: |