Atrial Evoked Response Integral for Automatic Capture Verification in Atrial Pacing
Autor: | Giuseppe Boriani, Gene A. Bornzin, Jeffrey D. Chatsworth Snell, Douglas Cameron, Mauro Biffi, Nils Holmström, Euljoon Park, Sergio Datteri |
---|---|
Rok vydání: | 2003 |
Předmět: |
Male
Pacemaker Artificial Preliminary analysis Electrocardiography Aged Cardiac Pacing Artificial Evoked Potentials Female Heart Atria Humans Atrial capture Medicine Lead (electronics) Intracardiac Electrogram Atrial pacing business.industry Pulse generator General Medicine Pacemaker Artificial Personal computer cardiovascular system Cardiac Pacing Cardiology and Cardiovascular Medicine business Biomedical engineering |
Zdroj: | Scopus-Elsevier |
ISSN: | 1540-8159 0147-8389 |
DOI: | 10.1046/j.1460-9592.2003.00026.x |
Popis: | BORIANI, G.,et al.:Atrial Evoked Response Integral for Automatic Capture Verification in Atrial Pacing. Beat-by-beat Autocapture is currently limited to operation in the ventricle with bipolar leads. The authors investigated the integral of the negative-going portion of the atrial evoked response integral (AERI) as a potential resource for verification of atrial capture. Intracardiac electrogram signals were collected from 59 patients (ages 67.8 ± 15.1 years) with bipolar, low polarization atrial leads. The signals were collected over a mean period of 6.1 months (minimum 4 days) after lead implantation. St. Jude Medical Affinity pulse generators were used to perform automatic capture threshold tests while the electrogram signals were recorded by a Model 3510 programming device. These signals were transferred to a personal computer in digital form for later analysis. The AERI was calculated at each programmable pacing voltage until capture was lost. The difference between the polarization integral at loss of capture and evoked response integral with successful capture was sufficient to justify enabling the atrial Autocapture feature in 53 of 59 patients in whom bipolar pacing and unipolar sensing was performed. The authors developed a calibration routine to identify automatically those patients in whom atrial Autocapture could be programmed On, based on the polarization integral at loss of capture, the estimated maximum polarization integral, and the AERI. Preliminary analysis indicated that the AERI is a practical resource for beat-by-beat atrial capture detection when used with low polarization leads. (PACE 2003; 26[Pt. II]:248–252) |
Databáze: | OpenAIRE |
Externí odkaz: |