Comparative in vitro mechanical evaluation of techniques using a 2.0 mm locking fixation system for simulated fractures of the mandibular body
Autor: | Andrezza Lauria de Moura, Renato Sawazaki, Roger William Fernandes Moreira, Raquel Correia de Medeiros |
---|---|
Rok vydání: | 2014 |
Předmět: |
Models
Anatomic medicine.medical_specialty Mandibular fracture Bone Screws computer.software_genre Fixation (surgical) Load testing Fracture Fixation Internal Mandibular body Mandibular Fractures Bone plate medicine Alloys Humans Mechanical Evaluation Orthodontics Titanium business.industry Neutral zone Equipment Design Mechanical resistance medicine.disease Surgery Biomechanical Phenomena Otorhinolaryngology Stress Mechanical Oral Surgery business computer Bone Plates Aluminum |
Zdroj: | Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery. 43(3) |
ISSN: | 1878-4119 |
Popis: | Purpose To perform a comparative evaluation of the mechanical resistance of simulated fractures of the mandibular body which were repaired using different fixation techniques with two different brands of 2.0 mm locking fixation systems. Materials and methods Four aluminum hemimandibles with linear sectioning simulating a mandibular body fracture were used as the substrates and were fixed using the two techniques and two different brands of fixation plate. These were divided into four groups: groups I and II were fixed with one four-hole plate, with four 6 mm screws in the tension zone and one four-hole plate, with four 10 mm screws in the compression zone; and groups III and IV were fixed with one four-hole plate with four 6 mm screws in the neutral zone. Fixation plates manufactured by Toride were used for groups I and III, and by Traumec for groups II and IV. The hemimandibles were submitted to vertical, linear load testing in an Instron 4411 servohydraulic mechanical testing unit, and the load/displacement (3 mm, 5 mm and 7 mm) and the peak loads were measured. Means and standard deviations were evaluated applying variance analysis with a significance level of 5%. Results The only significant difference between the brands was seen at displacements of 7 mm. Comparing the techniques, groups I and II showed higher mechanical strength than groups III and IV, as expected. Conclusion For the treatment of mandibular linear body fracture, two locking plates, one in the tension zone and another in the compression zone, have a greater mechanical strength than a single locking plate in the neutral zone. |
Databáze: | OpenAIRE |
Externí odkaz: |