Study of Image Quality(contrast) and Reduction of Patient Dose by Using Heavy Metal Filters

Autor: Ken-Ichi Yoshida, Taizou Sanada, Hiroko Nishimura, Yona Oishi, Hideyuki Iwanaga, Yuichi Sano, Akiyoshi Otsuka, Kazuo Fujimoto, Ken-ichiro Yasui
Rok vydání: 2002
Předmět:
Zdroj: Japanese Journal of Radiological Technology. 58:109-114
ISSN: 1881-4883
0369-4305
Popis: We studied image quality (contrast) and patient dose reduction using heavy metal filters in lumbar spine and abdomen x-ray examination. Heavy metal filters used in this study are gadolinium, holmium and ytterbium and these combinations. These filters have k-absorption edge in the range from 50 to 70 keV. Image quality and patient dose in 70-90 kV tube voltage with heavy metal filters were compared with 80 kV tube voltage without filter. Image quality was improved in four percent and patient dose could be reduced by 30%. However, tube loading increased from 1.6 to 2.2 times. It was found that the best filter choices gave better image and reduced patient dose compared to without filter.