Detection of in vitro proximal caries in storage phosphor plate radiographs scanned with different resolutions

Autor: W E R Berkhout, Gang Li, G.C.H. Sanderink, P.F. van der Stelt, M Martins
Přispěvatelé: Orale Radiologie (OUD, ACTA)
Rok vydání: 2008
Předmět:
Zdroj: Li, G, Berkhout, W E R, Sanderink, G C H, Martins, M & van der Stelt, P F 2008, ' Detection of in vitro proximal caries in storage phosphor plate radiographs scanned with different resolutions ', Dentomaxillofacial Radiology, vol. 37, no. 6, pp. 325-329 . https://doi.org/10.1259/dmfr/62591340
Dentomaxillofacial Radiology, 37(6), 325-329. British Institute of Radiology
ISSN: 0250-832X
DOI: 10.1259/dmfr/62591340
Popis: Objectives: To investigate the effect of the scanning resolution of storage phosphor plate (SPP) radiographs on the detection of proximal caries lesions. Methods: 10 dentists evaluated 72 proximal surfaces of premolars with respect to caries from SPP radiographs scanned with theoretical spatial resolutions of: (1) the Digora FMX at 7.8 lp mm-1; (2) the Digora Optime at both 7.8 lp mm-1 and 12.5 lp mm-1; and (3) the Dürr VistaScan at 10 lp mm-1 and 20 lp mm-1, respectively. The lesions were validated by histological examination. Receiver operating characteristic (ROC) analysis was employed. Results: The Az value for the radiographs scanned with the Dürr VistaScan at 10 lp mm-1 is significantly lower than those for the other series of radiographs (P = 0.000). Conclusions: For SPP radiographs, an increased theoretical spatial resolution per se is not related to an improved detection of proximal caries.