Intraperitoneal pressure: Stability over time and validation of Durand's measurement method
Autor: | Alicia Sobrino-Pérez, Cristina Barrios-Rebollo, Alfonso Pérez-Escudero, Lucila Fernández-Arroyo, Vicente Pérez-Díaz, Carmen Sánchez-Fonseca, Ana Dorado-García, Carmen Gutiérrez-Martín, Berta Martín-Alcón |
---|---|
Přispěvatelé: | Centre de Recherches sur la Cognition Animale (CRCA), Centre de Biologie Intégrative (CBI), Université Toulouse III - Paul Sabatier (UT3), Université Fédérale Toulouse Midi-Pyrénées-Université Fédérale Toulouse Midi-Pyrénées-Centre National de la Recherche Scientifique (CNRS)-Université Toulouse III - Paul Sabatier (UT3), Université Fédérale Toulouse Midi-Pyrénées-Université Fédérale Toulouse Midi-Pyrénées-Centre National de la Recherche Scientifique (CNRS)-Institut des sciences du cerveau de Toulouse. (ISCT), Université Toulouse - Jean Jaurès (UT2J)-Université Toulouse III - Paul Sabatier (UT3), Université Fédérale Toulouse Midi-Pyrénées-Université Fédérale Toulouse Midi-Pyrénées-CHU Toulouse [Toulouse]-Institut National de la Santé et de la Recherche Médicale (INSERM)-Centre National de la Recherche Scientifique (CNRS)-Université Toulouse - Jean Jaurès (UT2J)-CHU Toulouse [Toulouse]-Institut National de la Santé et de la Recherche Médicale (INSERM)-Centre National de la Recherche Scientifique (CNRS)-Centre National de la Recherche Scientifique (CNRS), Université Fédérale Toulouse Midi-Pyrénées-Université Fédérale Toulouse Midi-Pyrénées-Centre National de la Recherche Scientifique (CNRS) |
Rok vydání: | 2020 |
Předmět: |
2. Zero hunger
Adult Measurement method business.industry medicine.medical_treatment 030232 urology & nephrology Central venous pressure General Medicine 030204 cardiovascular system & hematology [SDV.MHEP.UN]Life Sciences [q-bio]/Human health and pathology/Urology and Nephrology Peritoneal dialysis Body Mass Index 03 medical and health sciences 0302 clinical medicine Nephrology medicine Pressure Humans Intraperitoneal pressure Peritoneum business Nuclear medicine Peritoneal Cavity Peritoneal Dialysis ComputingMilieux_MISCELLANEOUS |
Zdroj: | Peritoneal Dialysis International Peritoneal Dialysis International, Multimed Inc.; 1999, 2020, pp.089686082097312. ⟨10.1177/0896860820973120⟩ |
ISSN: | 1718-4304 0896-8608 |
DOI: | 10.1177/0896860820973120⟩ |
Popis: | Intraperitoneal pressure (IPP) is gaining consideration as a relevant parameter of peritoneal dialysis (PD) in adults, although many of its aspects are still pending clarification. We address here its stability over time and the validity of the usual method of clinical measurement, as proposed by Durand in 1992 but never specifically validated. We performed this validation by comparing Durand’s method and direct measurements with a central venous pressure system. We performed a total of 250 measurement pairs in 50 patients with different intraperitoneal volumes plus in-vitro measurements with a simulated peritoneum. Absolute differences between the two systems in vivo were 0.87 ± 0.91 cmH2O (range 0–5 cmH2O); only 6.4% of them were ≥3 cmH2O. In vitro results for both methods were identical. We also compared IPP measurements in the same patient separated by 1–4 h (514 measurement pairs in 136 patients), 1 week (92 pairs in 92 patients), and 2 years (34 pairs in 17 patients). Net differences of measurements separated by hours or 1 week were close to 0 cmH2O, with oscillations of 1.5 cmH2O in hours and 2.3 cmH2O in 1 week. IPP measured 2 years apart presented a net decrease of 2.5 ± 4.9 cmH2O, without correlation with body mass index changes or any other usual parameter of PD. In hours, 7% of IPP differences were >3 cmH2O, 22% in 1 week, and 50% in 2 years. In conclusion, Durand’s method is precise enough to measure IPP in peritoneal dialysis. This parameter is not stable over long timescales, so it is necessary to use recent measurements. |