Comparison of conventional versus customised Eurosil-4 Pink bolus for radiotherapy of the chest wall

Autor: Ashlesha Gill, Warwick Smith, Andrew Hirst, Mahsheed Sabet, Zaid Alkhatib, Suki Gill, Pejman Rowshanfarzad
Rok vydání: 2021
Předmět:
Zdroj: PloS one. 17(5)
ISSN: 1932-6203
Popis: Introduction In radiotherapy, the presence of air gaps near a tumour can lead to underdose to the tumour. In this study, the impact of air gaps on dose to the surface was evaluated. 3D-printing was used to construct a Eurosil-4 Pink bolus customised to the patient and its dosimetric properties were compared with that of Paraffin wax bolus. Methods Surface dose was measured for flat sheets of Eurosil-4 Pink bolus with different thicknesses. Different air gap thicknesses were inserted between the bolus and the surface, and dose was measured for each air gap using 10 cm × 10 cm fields. This was repeated with the effective field size calculated from the patient plan. Surface dose was measured for varying angles of incidence. A customised chest phantom was used to compare dose for two customised Eurosil-4 Pink boluses, and commonly used Paraffin wax bolus. Results The surface dose was found to be highest for 1.1 cm thick bolus. The decrease in surface dose for the Eurosil-4 Pink bolus was minimal for the 10 cm × 10 cm field, but higher for the effective field size and larger angles of incidence. For instance, the dose was reduced by 6.2% as a result of 1 cm air gap for the effective field size and 60 degree angle of incidence. The doses measured using Gafchromic film under the customised Eurosil-4 Pink boluses were similar to that of the Paraffin wax bolus, and higher than prescribed dose. Conclusions The impact of air gaps can be significant for small field sizes and oblique beams. A customised Eurosil-4 Pink bolus has promising physical and dosimetric properties to ensure sufficient dose to the tumour, even for treatments where larger impact of air gaps is suspected.
Databáze: OpenAIRE