Autor: |
C. Trosseille, M. Burillo, P. Stemmler, G. Oudot, J. P. Fronty, G. Soullié, Catherine Goulmy, Patrick Brunel, C. Zuber, D. Gontier, I. Moreau, C. Rubbelynck, S. Bazzoli |
Rok vydání: |
2016 |
Předmět: |
|
Zdroj: |
The Review of scientific instruments. 87(11) |
ISSN: |
1089-7623 |
Popis: |
A prototype of a picosecond x-ray streak camera has been developed and tested by Commissariat a l’Energie Atomique et aux Energies Alternatives to provide plasma-diagnostic support for the Laser Megajoule. We report on the measured performance of this streak camera, which almost fulfills the requirements: 50-μm spatial resolution over a 15-mm field in the photocathode plane, 17-ps temporal resolution in a 2-ns timebase, a detection threshold lower than 625 nJ/cm2 in the 0.05–15 keV spectral range, and a dynamic range greater than 100. |
Databáze: |
OpenAIRE |