Improved pressurized Marinelli beaker measurements of radioactive xenon in air

Autor: Paul Fabian, Matthew G. Watrous, Tracy P. Houghton, Troy A. Robinson, Francisco Fernández, Nick R. Mann, John W. Peterson, Mark Reavis, Pat Hipp
Rok vydání: 2017
Předmět:
Zdroj: Applied Radiation and Isotopes. 126:20-22
ISSN: 0969-8043
DOI: 10.1016/j.apradiso.2016.12.042
Popis: INL has shown that a Marinelli beaker geometry can be used for the measurement of radioactive xenon in air using an aluminum Marinelli. A carbon fiber Marinelli was designed and constructed to improve overall performance. This composite Marinelli can withstand sample pressures of 276bar and achieve approximately a 4x performance improvement in the minimum detectable concentrations (MDCs) and concentration uncertainties. The MDCs obtained during a 24h assay for 133Xe, 131mXe, and 135Xe are: 1.4, 13, and 0.35Bq/m3.