Development of a quasi‐humanoid phantom to perform dosimetric and radiobiological measurements for out‐of‐field doses from external beam radiation therapy

Autor: Marta, Kruszyna-Mochalska, Agnieszka, Skrobala, Piotr, Romanski, Adam, Ryczkowski, Wiktoria, Suchorska, Katarzyna, Kulcenty, Igor, Piotrowski, Dorota, Borowicz, Natalia, Matuszak, Julian, Malicki
Rok vydání: 2022
Předmět:
Zdroj: Journal of Applied Clinical Medical Physics. 23
ISSN: 1526-9914
Popis: Our understanding of low dose, out-of-field radiation and their radiobiological effects are limited, in part due to the rapid technological advances in external beam radiotherapy, especially for non-coplanar and dynamic techniques. Reliable comparisons of out-of-field doses produced by advanced radiotherapy techniques are difficult due to the limitations of commercially available phantoms. There is a clear need for a functional phantom to accurately measure the dosimetric and radiobiological characteristics of out-of-field doses, which would in turn allow clinicians and medical physicists to optimize treatment parameters. We designed, manufactured, and tested the performance of a quasi-humanoid (Q-H) adult phantom. To test the physics parameters, we used computed tomography (CT) scans of assembled Q-H phantom. Static open field and dynamic techniques were measured both in- and out-of-field with ionization chambers and radiochromic films for two configurations (full solid and with water-filled containers). In the areas simulating soft tissues, lung, and bones, median Hounsfield units and densities were, respectively: 129.8, -738.7, 920.8 HU and 1.110, 0.215, 1.669 g/cm
Databáze: OpenAIRE